Analytical system Fully automated for routine, emergency test, urine and homogeneous immunoassays.
Analytical methods Clinical chemistry and immunological parameters. End point, kinetic assays, fixed time kinetics, and ISE option.
Simultaneously processed analytes Up to 96 photometric tests and 3 ISE.
Throughput AU5420: Up to 3200 photometric tests/hour; ISE up to 600 samples/hour. AU5430: Up to 4800 photometric tests/hour; ISE up to 600 samples/hour.
Sample feeder Racks with 10 samples each (bar-codes on primary tubes and on racks); capability of 300 samples; continuous loading.
Sample tubes In primary and secondary tubes diameter from 11.5 to 16 mm, height from 55 to 102 mm.
Sample volume 1.6 to 25 µl in 0.1 µl steps (1 µl for repeats).
Reagent supply Refrigerated (4 °C - 12 °C); 48 positions for R1; 48 positions for R2/module.
Reagent volume 1. Reagent 15-250 µl; 2. Reagent 15-250 µl.
Total reaction volume 120-430 µl.
Reaction cuvette Quartz cuvettes.
Reaction time Up to 8 minutes, 22 seconds.
Reaction temperature 37 °C.
Mixing method With rotating paddles after dispensing sample and reagent.
Photometry system Direct assay through the reaction cuvette (0-2.5 OD); mono-, bichromatic. measurements possible
Wavelength 13 different wavelengths between 340 and 800 nm.
Cuvette cleaning Comprehensive cleaning with detergents.
Calibration Linear and non-linear (linear interpolation), refrigerated calibrator positions.
Data processing Calculated parameters, blank value correction and correlation. Definition of normal value range, measuring range, pathological values and repeats. Quality controls, printout of abnormal values, pending sample work list and repeat run sample list.
Test requisition Individual and profile test requisition via on-line, mouse, function keys or optional touch screen.
Safety Clot detection and crash prevention for sample and reagent dispenser.
On-line Full uni- and bi-directional communication possible.
Software Windows NT®; modem capability.
Dimensions (W x H x D) mm AU5420 ANL: 3960 x 1260 x 1140 DPR: 700 x 1430 x 700mm AU5430 ANL: 5090 x 1260 x 1140 DPR: 700 x 1430 x 700mm

Specifications and design are subject to change without notice.