| Analytical system |
Fully automated, random access, STAT capability |
| Analytical methods |
Heterogeneous immunoassays Solid phase: dispensed magnetic particles Detection: chemiluminescence caused by hydrolysis of AMPPD (1) Catalyst: Alkaline Phosphatase |
| Simultaneously processed analytes |
Up to 24 |
| Throughput: |
240 tests per hour |
| Sample feeder : |
Continuous loading Capacity of 300 samples Olympus racks with 10 samples each, compatible to AU Clinical Chemistry analysers PSID (bar codes on primary tubes and racks) |
| Sample tubes: |
All leading commercially available tubes Outer diameter from 11.5 to 16 mm Length from 55 to 102 mm Micro cups |
| Sample volume: |
10µl - 100µl |
| Reagent supply : |
Two reagent refrigerators (4 °C - 12 °C) |
| Reaction temperature: |
Immunochemical reaction: 37 °C ± 0.1 °C |
| Light emitting reaction: |
37 °C ± 0.7 °C |
| Calibration: |
Master calibration set by CD together with assay protocol Analyser calibration by measurement of one or two calibration materials |
| Test requisition : |
Individual and profile test requisition via on-line, mouse, function keys or optional touch screen |
| Sample integrity : |
Clot detection, level detection; crash prevention |
| On-line connectivity : |
Bidirectional communication possible |
Dimensions: (W x H x D)/mm |
1700 x 1200 x 1440 |
| Mass : |
710 kg |
| Power supply : |
210 V; 220 V; 230 V; 240 V / 50 Hz; 60 Hz / 4.4 kVA |
| (1) AMPPD: 4-methoxy-4-(3-phosphatephenyl 1) spiro{1,2-dioxetane-3,2'-damantane}, disodium salt |